Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Dual wavelength analysis"

From Bioblast
Line 1: Line 1:
{{MitoPedia
{{MitoPedia
|abbr=n.a.
|description=If a sample contains a number of absorbing substances, it is sometimes possible to select discrete pairs of wavelengths at which the change in [[absorbance]] of a particular substance (due to oxidation or reduction, for example) is largely independent of changes in the [[absorbance]] of other substances present. '''Dual wavelength analysis''' can be carried out for [[cytochrome c]] by subtracting the [[absorbance]] at 540 nm from that at 550nm in order to give a measure of the degree of reduction. Similarly, by subtracting the [[absorbance]] at 465 nm from that at 444 nm, an indicator of the [[redox state]] of [[Complex IV | cytochrome ''aa''<sub>3</sub>]] can be obtained.
|description=If a sample contains a number of absorbing substances, it is sometimes possible to select discrete pairs of wavelengths at which the change in [[absorbance]] of a particular substance (due to oxidation or reduction, for example) is largely independent of changes in the [[absorbance]] of other substances present. '''Dual wavelength analysis''' can be carried out for [[cytochrome c]] by subtracting the [[absorbance]] at 540 nm from that at 550nm in order to give a measure of the degree of reduction. Similarly, by subtracting the [[absorbance]] at 465 nm from that at 444 nm, an indicator of the [[redox state]] of [[Complex IV | cytochrome ''aa''<sub>3</sub>]] can be obtained.
}}
}}
{{MitoPedia concepts}}
{{MitoPedia methods
{{MitoPedia methods
|mitopedia method=Spectrophotometry
|mitopedia method=Spectrophotometry
}}
}}
{{MitoPedia SUIT}}
{{MitoPedia topics}}
{{MitoPedia topics}}

Revision as of 23:27, 27 January 2016


high-resolution terminology - matching measurements at high-resolution


Dual wavelength analysis

Description

If a sample contains a number of absorbing substances, it is sometimes possible to select discrete pairs of wavelengths at which the change in absorbance of a particular substance (due to oxidation or reduction, for example) is largely independent of changes in the absorbance of other substances present. Dual wavelength analysis can be carried out for cytochrome c by subtracting the absorbance at 540 nm from that at 550nm in order to give a measure of the degree of reduction. Similarly, by subtracting the absorbance at 465 nm from that at 444 nm, an indicator of the redox state of cytochrome aa3 can be obtained.



MitoPedia methods: Spectrophotometry 

Template:MitoPedia SUIT