MiPNet08.16 pH calibration
Laboratory protocol: pH measurement and temperature dependence of pH. Β»Bioblast pdfΒ« |
Β» Versions
OROBOROS (2014-03-06) Mitochondr Physiol Network
Abstract: Gnaiger E (2014) Laboratory protocol: pH measurement and temperature dependence of pH. Mitochondr Physiol Network 08.16(07): 1-4.
pH of blood and intracellular compartments is tightly regulated. Blood pH at 37 Β°C is 7.4. However, when temperature is lowered, the pH of blood and of intracellular buffer systems increases, which has been reported as early as 1927 but was ignored for the following 30 years. The phenomenon of a change in blood pH with temperature of about -0.016 U/Β°C was rediscovered by comparative physiologists in studies of "cold blooded" vertebrates (turtles, fish), and is was recognized soon that the same pH/temperature relation applies to "warm blooded" mammals (rat, human). The physicochemical basis of pH/temperature relations and the consequences for acid-base balance and protein function were primarily analyzed by Rahn and Reeves (1979). Besides the importance for physiological and biochemical systems, the temperature dependence of pH of buffer systems has experimental significance for the measurement of pH at different temperatures, and for the choice of buffer systems when designing experiments at various temperatures.
- >> Product: OROBOROS O2k, O2k-Catalogue
β’ Keywords: HRR-MultiSensor
β’ O2k-Network Lab: AT_Innsbruck_OROBOROS
Labels: MiParea: Instruments;methods
Regulation: pH, Temperature
HRR: pH, Protocol"Protocol" is not in the list (Oxygraph-2k, TIP2k, O2k-Fluorometer, pH, NO, TPP, Ca, O2k-Spectrophotometer, O2k-Manual, O2k-Protocol, ...) of allowed values for the "Instrument and method" property.
O2k-SOP
Further information
In DatLab, the two-point calibration calculations are performed automatically (equations in MiPNet12.08).