Polarization voltage: Difference between revisions

From Bioblast
No edit summary
(added technical support banner)
 
(9 intermediate revisions by 3 users not shown)
Line 1: Line 1:
{{MitoPedia
{{Technical support}}
{{MitoPedia without banner
|abbr=''U''
|abbr=''U''
|description=A '''polarization voltage''' of 600 mV to 800 mV is applied between anode and cathode of the [[polarographic oxygen sensor]], resulting in a current when oxygen is consumed. The current is converted by the electronics to a voltage (raw signal) which must not be confused with the polarization voltage.
|description=A '''polarization voltage''' of 600 mV to 800 mV is applied between anode and cathode of the [[polarographic oxygen sensor]], resulting in a current when oxygen is consumed. The current is converted by the electronics to a voltage (raw signal) which must not be confused with the polarization voltage.
|info=[[MiPNet22.11 O2k-FluoRespirometer manual]]
}}
}}
{{MitoPedia concepts}}
{{MitoPedia methods}}
{{MitoPedia O2k and high-resolution respirometry
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=O2k hardware, DatLab
|mitopedia O2k and high-resolution respirometry=O2k hardware, DatLab
}}
}}
{{MitoPedia topics}}
{{Technical support integrated}}
 
== DatLab ==
== DatLab ==
The polarization voltage is set in [[DatLab]] in the "Oxygraph-2k" \ "O2k-Control" window ([[MiPNet19.18A O2k-Start]]). The default value is 800 mV.
:::: The polarization voltage is set in [[DatLab]] in the [[O2k control]] window [F7] ([[MiPNet22.11 O2k-FluoRespirometer manual]]). The default value is 800 mV.
 
 
 
[[Category:Technical service]]

Latest revision as of 09:20, 5 April 2022


                  


O2k-Open Support

Polarization voltage



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

A polarization voltage of 600 mV to 800 mV is applied between anode and cathode of the polarographic oxygen sensor, resulting in a current when oxygen is consumed. The current is converted by the electronics to a voltage (raw signal) which must not be confused with the polarization voltage.

Abbreviation: U

Reference: MiPNet22.11 O2k-FluoRespirometer manual


MitoPedia O2k and high-resolution respirometry: O2k hardware, DatLab 


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



DatLab

The polarization voltage is set in DatLab in the O2k control window [F7] (MiPNet22.11 O2k-FluoRespirometer manual). The default value is 800 mV.