Did you know that the proper use of TIP2k-Filter Papers may avoid an unstable O2 slope neg. signal during Instrumental background correction?
- »Instrumental background« and »TIP2k Filter Papers«
- See the scheme:
Bioblast alert 2020
Prep for circular done here: Talk:Bioblast alert 2020
MitoPedia alert
Next
- Jun-14 Th
How can I measure ATP production in cells?
Respiration measurements with living cells provides information on parameters such as free ROUTINE activity. MitoPedia may help to clarify why defining it as ATP production is inappropriate
- »SUIT protocols to measure ATP production« Learn more about how to measure ATP production in mitochondrial preparations with Magnesium Green -
- »Magnesium Green«
Already added to Agenda - Next and to be added to Mitopedia alert
O2k-Publications alert
Next
Already added to Agenda - Next and to be added to O2k-Publications alert
O2k-Open Support alert
Next
- While in Home Office: Gnaigere (talk) 14:03, 19 March 2020 (CET) https://wiki.oroboros.at/index.php/Talk:O2k-Open_Support_alert - Many labs will move to home office mode. In line with our special DL7.4 home office offer (see Agenda today), therefore, I suggest to post in our Agenda analysis-linked O2k-Open Support cases during this time, and collect real-time operation cases for future releases.
- Jun-01 Mo
Did you know that the proper use of TIP2k-Filter Papers may avoid an unstable O2 slope neg. signal during Instrumental background correction?
- »Instrumental background« and »TIP2k Filter Papers «
Already added to Agenda - Next and to be added to O2k-Open support alert
- May-18 Mo
Do you know that a humid O2k-environment and/or liquid spillage may lead to a high raw POS signal [V]?
See how to clean and dry the electrical connections of the OroboPOS-Connector.
- »Cleaning the electrical connections«
- May-25 Mo
Do you know how the "switch components" approach can help you find what is causing your O2k to not reach the quality control specifications?
Do not despair if specifications given in the O2k-Manuals are not obtained. See how to troubleshoot on your own.
- »Switch components approach«
Removed for correction
Do you perform the leaky chamber test after chamber assembly?
An unnoticed leak from the chamber may affect your data, therefore we recommend performing the Leaky chamber test - especially after chamber reassembly.
- »Leaky chamber test« and » https://intern.oroboros.at/index.php/Overnight_test_after_O2k-chamber_assembly«
O2k-Open Support alert circulars
O2k-Open Support alert circular 2020(01): in prep
New A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.
See how you can check that the two blue OroboPOS-Holders are intact.
- »O2k-OroboPOS-Holder«
New Calibration parameters are transferred to experimental DatLab files
Save calibration information from selected Marks while DatLab is connected to the O2k during an instrumental calibration run. Air calibration and zero oxygen calibration can be performed and saved in a single DatLab file or in separate DatLab files.
- » Oxygen calibration - DatLab«
Do you perform O2k-Chamber volume calibration?
The volume calibration is essential to ensure that the chamber volume is correctly calibrated to its standard volume of 2 mL (or 0.5 mL for the O2k-sV-Module).
- »O2k-Chamber MitoPedia«
Do you use the correct O2 solubility factor (FM) in the ‘O2 calibration’ window in DatLab?
The O2 solubility factor varies according to the medium used and influences the air calibration.
- »Oxygen solubility factor«
In preparation
For discussion
Here we discuss suggestions for the Oroboros Ecosystem Agenda 2020
width="110" align="center" Controlled stability of the MitoKit-CII based on HPLC and LCMS analysis: No degradation of MitoKit-CII compounds MitoKit-CII/Succinate-nv (NV118) and MitoKit-CII/Malonate-nv (NV161) for following storage conditions: - Proven -80 °C solid stability for more than 3 years (3 years 7 months)
- Proven -20 °C solid stability for 1 year
- Proven DMSO stability at room temperature for 1 week
- Proven DMSO stability at -20 °C for 6 months (ongoing)
As part of the H2020 SME NextGen-O2k project we are developing a PhotoBiology module that will allow to study the effect of different light intensities on O2 consumption and production
-»PhotoBiology«
Do you perform a stirrer test during oxygen calibration?
In DatLab 7.4, instrumental DL-Protocols guide you automatically towards performing the stirrer test in the process of air calibration. It shows immediately, if your OroboPOS has a satisfactory response time. Further details in
- » MitoPedia: »Oxygen calibration - DatLab
Is the noise of the ‘O2 slope neg’ acceptable?
If you are not certain that the noise of the ‘O2 slope neg’ is within a suitable range or you want to confirm that your experimental data will not be affected, see:
- » MiPNet06.03 POS-calibration-SOP