  • Recommendation: Set "Slope smoothing" to 20 for performing and analyzing experiments with biological sample (strongly recommended in the case of uncoupler titrations). A slope smoothing of 20 increases the time or wavelength resolution, but increases the noise of the signal.
The following DatLab traces illustrate an example of "Slope smoothing" using 20 versus 40 data points:

Amp slope smoothing 20.png

Amp slope smoothing 40.png