== Apr 2021 ==
== Apr 2021 ==
*'''Apr-27 Tu'''
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Yellow-take care.png|25px|Take care]]
|''' Why is the fluorescence signal constantly at 10 V or at 0 V?'''<br>
Follow these steps to localize and solve the problem:
<br>
- ยป'''[[Troubleshooting:_fluorescence_module#Fluorescence_signal_is_constantly_at_10_V|Fluorescence signal is constantly at 10 V]]'''ยซ and ยป'''[[Defective_fluorescence_module#Zero_fluorescence_signal|0 V fluorescence signal]]'''ยซ
|}
----
*'''Apr-20 Tu''' ย 
*'''Apr-20 Tu''' ย 
:{|
:{|

Revision as of 08:04, 26 April 2021

Currently Answered Questions from the ยปOroboros Ecosystem Agendaยซ

Apr 2021

  • Apr-27 Tu
O2k-Open SupportTake care Why is the fluorescence signal constantly at 10 V or at 0 V?

Follow these steps to localize and solve the problem:
- ยปFluorescence signal is constantly at 10 Vยซ and ยป0 V fluorescence signalยซ


  • Apr-20 Tu
O2k-Open SupportWarning: loss of data Noisy fluorescence signal?

Follow these steps to localize and solve the problem:
- ยปNoisy fluorescence signalยซ


  • Apr-13 Tu
O2k-Open SupportWarning: loss of data Do you know what to do when the fluorescence signal starts at a negative value?

A negative signal may be caused by wet Fluo-Sensors. For further information:
- ยปNegative fluorescence signalยซ


  • Apr-06 Tu
O2k-Open Support Warning: loss of data A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.

See how you can check that the OroboPOS-Holders are intact.
- ยปO2k-OroboPOS-Holderยซ


Mar 2021

  • Mar-25 Th
O2k-Open SupportInfo: improve O2k-checklist: Get started with an O2k-experiment

The O2k-checklist provides an overview of all essential steps for starting an O2k-experiment.
- ยป MiPNet20.04_O2k-checklistยซ 
- ยป communicated by ยปCecatto Cristianeยซ and ยปSchmitt Sabineยซ


Feb 2021

  • Feb-17 We
O2k-Open SupportInfo: improve Do you know how to prepare permeabilized muscle fibers?

The application of high-resolution respirometry in permeabilized muscle fibers provides a sensitive diagnostic tool for the analysis of mitochondrial (dys)function in small biopsy samples.
- ยปO2k-Videosupport: Fiber permeabilizationยซ 
- ยปDoerrier 2018 Methods Mol Biolยซ 
- communicated by ยปDoerrier Carolinaยซ and ยปTindle-Solomon Lisaยซ


  • Feb-08 Mo
O2k-Open SupportInfo: improve Do you know how to properly insert the ISS-tip into the O2k-chamber?

With the stirrer bar still rotating, insert the ISS-tip to the bottom of the O2k-Chamber, so that it points away from you.
- ยปMiPNet19.03 O2k-cleaning and ISS: Section 2ยซ 
- communicated by ยปDoerrier Carolinaยซ, ยปCecatto Cristianeยซ and ยปKomlodi Timeaยซ


  • Feb-02 Tu
O2k-Open SupportInfo: improve DatLab error messages: What do they mean and how to solve them?

- ยป DatLab error messagesยซ 
- communicated by ยปGarcia-Souza Luiz Fยซ and ยปTindle-Solomon Lisaยซ


Jan 2021

  • Jan-11 Mo
O2k-Open SupportInfo: improve Are you a beginner or need a refresher on using SUIT protocols with DatLab7.4?

- ยป O2k-Videosupport: How to perform an experiment with a SUIT DL-Protocolยซ and ยป Run DL-Protocol/Set O2 limit in DatLab ยซ 
- communicated by ยปKomlodi Timeaยซ and Schmitt Sabineยซ