== Instability of the oxygen signal in the open O2k-Chamber == | |||
ย | |||
In an [[open chamber]] of the O2k the liquid phase in the chamber (aqueous medium) is in equilibrium with the atmosphere. All oxygen consumed by the polarographic oxygen sensor ([[POS]]) is immediately replaced from the atmosphere. The oxygen signal therefore has to be constant and the (negative) time derivative of the oxygen signal, called "O2 slope uncorr." in DatLab, has to be zero. The background corrected oxygen flux is meaningless for the open chamber situation. This is because the background correction at air saturation subtracts the consumption of oxygen by the sensor from the negative slope, when diffusion into and out of the chamber is zero at air saturation. Therefore, the background-corrected oxygen flux in the open chamber at air saturation is shown as a negative value. To avoid this apparent artefact, the "O2 slope uncorr." is selected to be shown while the chamber is open. Only Graph Layouts that display "O2 slope uncorr." are suitable for assessing the stability of the oxygen signal when the chamber is open. Such Layouts are: ย | In an [[open chamber]] of the O2k the liquid phase in the chamber (aqueous medium) is in equilibrium with the atmosphere. All oxygen consumed by the polarographic oxygen sensor ([[POS]]) is immediately replaced from the atmosphere. The oxygen signal therefore has to be constant and the (negative) time derivative of the oxygen signal, called "O2 slope uncorr." in DatLab, has to be zero. The background corrected oxygen flux is meaningless for the open chamber situation. This is because the background correction at air saturation subtracts the consumption of oxygen by the sensor from the negative slope, when diffusion into and out of the chamber is zero at air saturation. Therefore, the background-corrected oxygen flux in the open chamber at air saturation is shown as a negative value. To avoid this apparent artefact, the "O2 slope uncorr." is selected to be shown while the chamber is open. Only Graph Layouts that display "O2 slope uncorr." are suitable for assessing the stability of the oxygen signal when the chamber is open. Such Layouts are: ย | ||
* 01 Calibration Exp Gr3-Temp | * 01 Calibration Exp Gr3-Temp | ||
Line 38: | Line 42: | ||
'''Solutions''' | '''Solutions''' | ||
:* The oxygen level in the gas phase may not correspond to air. Move the stopper upwards to exchange the gas phase and insert it partially using the [[Stopper-Spacer]] as a guide. | :* The oxygen level in the gas phase may not correspond to air. Move the stopper upwards to exchange the gas phase and insert it partially using the [[Stopper-Spacer]] as a guide. | ||
'''See also''' | '''See also''' | ||
:ยป [[ | :ยป [[O2k-technical support and open innovation]] | ||
ย | |||
[[Category:Technical service]] | [[Category:Technical service]] |
Revision as of 22:31, 7 February 2016
Description
Oxygen flux, JO2, is a specific quantity. Oxygen flux is oxygen flow, IO2 [molยทs-1 per system], divided by system size. Flux may be volume-specific (flow per volume [pmolยทs-1ยทml-1]), mass-specific (flow per mass [pmolยทs-1ยทmg-1]), or marker-specific (e.g. flow per mtDNA). Oxygen flux (e.g. per body mass, or per cell mass) is distinguished from oxygen flow (per subject, or per million cells).
Abbreviation: JO2
Reference: Gnaiger 2014 MitoPathways, Gnaiger 1993 PAC, Renner_2003_Biochim_Biophys_Acta
MitoPedia concepts: "MitoFit Quality Control System" is not in the list (MiP concept, Respiratory state, Respiratory control ratio, SUIT concept, SUIT protocol, SUIT A, SUIT B, SUIT C, SUIT state, Recommended, ...) of allowed values for the "MitoPedia concept" property.
MitoFit Quality Control System"MitoFit Quality Control System" is not in the list (Enzyme, Medium, Inhibitor, Substrate and metabolite, Uncoupler, Sample preparation, Permeabilization agent, EAGLE, MitoGlobal Organizations, MitoGlobal Centres, ...) of allowed values for the "MitoPedia topic" property.
MitoPedia methods:
Respirometry, "DatLab" is not in the list (Respirometry, Fluorometry, Spectrophotometry) of allowed values for the "MitoPedia method" property.
DatLab
Labels:
HRR: Theory
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Instability of the oxygen signal in the open O2k-Chamber
In an open chamber of the O2k the liquid phase in the chamber (aqueous medium) is in equilibrium with the atmosphere. All oxygen consumed by the polarographic oxygen sensor (POS) is immediately replaced from the atmosphere. The oxygen signal therefore has to be constant and the (negative) time derivative of the oxygen signal, called "O2 slope uncorr." in DatLab, has to be zero. The background corrected oxygen flux is meaningless for the open chamber situation. This is because the background correction at air saturation subtracts the consumption of oxygen by the sensor from the negative slope, when diffusion into and out of the chamber is zero at air saturation. Therefore, the background-corrected oxygen flux in the open chamber at air saturation is shown as a negative value. To avoid this apparent artefact, the "O2 slope uncorr." is selected to be shown while the chamber is open. Only Graph Layouts that display "O2 slope uncorr." are suitable for assessing the stability of the oxygen signal when the chamber is open. Such Layouts are:
- 01 Calibration Exp Gr3-Temp
- 02 Background Experiment
- 04 Flux per Volume uncorrected
- Z Trouble Shooting
The observation of a zero flux with an open chamber is an important performance parameter. It indicates that thermal stability and equilibrium of oxygen between the gas and aqueous phases have been reached. Therefore no experiment should be started before a zero "O2 slope uncorr." has been reached with an open chamber. The suggested criterion for signal stability is a "O2 slope uncorr." between -1 pmol/(s ml) and + 1 pmol/(s ml).
Problem
- The "O2 slope uncorr." does not reach the interval +/- 1 pmol/(s ml) even after a prolonged time (1 to 2 hours).
Solutions
- The oxygen level in the gas phase may not correspond to air. Move the stopper upwards to exchange the gas phase and insert it partially using the Stopper-Spacer as a guide.