Life Technologies, R302, 25 mg, store at -20 ยฐC, FW=380.83
Life Technologies, R302, 25 mg, store at -20 ยฐC, FW=380.83
:::'''Preparation of 5 mM storage solution (dissolved in ethanol)'''
:::'''Preparation of 5 mM storage solution (dissolved in ethanol)'''

Revision as of 12:52, 17 March 2019


high-resolution terminology - matching measurements at high-resolution


Rhodamine 123

Description

Rhodamine 123 is an extrinsic fluorophore and can be used as a probe to determine changes in mitochondrial membrane potential. Rh123 is a lipophilic cation that is accumulated by mitochondria in proportion to ฮ”ฯˆmt. Using ethanol as the solvent, the excitation maximum is511 nm and the emission maximum is 534 nm.The recommended excitation and emission wavelengths in PBS are 488 and 515-575 nm, respectively (Sigma-Aldrich).

Abbreviation: Rh123

Template:Keywords Force and membrane potential

Application in HRR

Life Technologies, R302, 25 mg, store at -20 ยฐC, FW=380.83

Preparation of 5 mM storage solution (dissolved in ethanol)
  1. Weight 1.9 mg of Rh123 and dissolve in 1 mL ethanol;
  2. Divide into 0.1 mL portions (use dark vials);
  3. Store at -20 ยฐC.
Preparation of 0.2 mM stock solution
  1. Dilute a 100 ยตL aliquot 5 mM storage solution with 2400 ยตL ethanol to reach 2.5 mL final volume;
  2. Divide into 0.2 mL portions (use dark vials);
  3. Store at -20 ยฐC.

Signal and output

  1. Signal: The O2k-Fluo LED2-Module is operated through the Amperometric (Amp)-Channel of the O2k, with electric current (ampere [A]) as the primary signal.
  2. Output: The focus of the output with Rh123 is on Type C: Force, mt-membrane potential.

Rh123 with the O2k-FluoRespirometer: exploratory experiments

ImtB Rhodamine 123 S-Pathway.png


MitoPedia methods: Fluorometry 


Labels: MiParea: Respiration, Instruments;methods 





HRR: Oxygraph-2k, O2k-Fluorometer, O2k-Protocol 

O2k-MultiSensor 